Page 16

background image

Operation
Eppendorf

®

μPlate G0.5

English (EN)

16

5.2.7

Loading samples

1. Open the μPlate G0.5.

2. Pipette the samples into the wells.

Abb. 5-7:μPlate G0.5 without pipetting aid

Fig. 5-7:

μPlate G0.5 without pipetting aid

3. Close the μPlate G0.5 lid immediately after loading.

4. Insert the μPlate G0.5 in the PlateReader AF2200.

The filled μPlate G0.5 should always be measured within 5 minutes in order to
prevent evaporation of the sample or the blank - and avoid imprecise results.

Use the pipetting aid if you need assistance with precise application.