A Stepdos pump is a microprocessor-controlled membrane dosing pump for liquids. It is used for

the quasi-continuous and consistently neutral or aggressive dosing of liquid (depending on the

pump model).
The pumps have two different operating modes:

o

Run mode

Operating function for volume flow rate dosing

o

Dispense mode

Operating function for infinitely variable batch dosing

&RQWURO

In run mode

1. Volume flow rate [ml/min]
2. Calibrator input [g]
5. Start / Stop

In dispense mode

3. Dispense volume [ml]
4. Dispense time [s]
5. Start / Stop

0HDVXUHPHQWV

In run mode

1. Volume flow rate [ml/min]

In dispense mode

2. Dispense volume [ml]


Automatic calibration is activated when scales are connected to the calibrator input (2). You can

specify how often calibration should take place per hour in the input fields. During calibration the set

volume flow rate is determined with the actual change in weight on the scales and the pump is

automatically readjusted.

1RWH

Please see separate operating instructions for more details of appliance properties.