A Stepdos pump is a microprocessor-controlled membrane dosing pump for liquids. It is used for
the quasi-continuous and consistently neutral or aggressive dosing of liquid (depending on the
pump model).
The pumps have two different operating modes:
o
Run mode
Operating function for volume flow rate dosing
o
Dispense mode
Operating function for infinitely variable batch dosing
&RQWURO
In run mode
1. Volume flow rate [ml/min]
2. Calibrator input [g]
5. Start / Stop
In dispense mode
3. Dispense volume [ml]
4. Dispense time [s]
5. Start / Stop
0HDVXUHPHQWV
In run mode
1. Volume flow rate [ml/min]
In dispense mode
2. Dispense volume [ml]
Automatic calibration is activated when scales are connected to the calibrator input (2). You can
specify how often calibration should take place per hour in the input fields. During calibration the set
volume flow rate is determined with the actual change in weight on the scales and the pump is
automatically readjusted.
1RWH
Please see separate operating instructions for more details of appliance properties.
To ensure correct communication, you must set the transfer parameters in the appliance’s serial
interface to the following values:
EDXG VWDUW ELW GDWD ELWV VWRS ELW SDULW\ QRQH 56 FRPPXQLFDWLRQ