beautypg.com

WIKA UT-11 User Manual

Page 21

background image

P

u

bl

is

he

d 1

6

.2.

04

C

Universal Pressure Transmitter UniTrans

Operation of Transmitters without Display

Subject to change due to technical modifications.

© Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD

WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiegand-Str. · 63911 Klingenberg ·

(09372) 132 - 710 · Fax - 706 · E-mail: [email protected] · www.wika.de

60

5.4

Calibration without Pressure

Determine the current reference values for the zero point and the span to be entered
in the transmitter before calibration. This is done as follows:

5.4.1

Zero Point Calibration

• Determine the hydrostatic pressure of the liquid’s surface that meets the zero point.

• Adjust this pressure in proportion to the sensor’s nominal pressure range.

• Multiply this proportion by 16 mA and add 4 mA to the result.

This produces the calculated current (value I

calc

), which is entered in the transmitter

and used to programm the zero point (0%).

Example:

A pressure transducer with 0 ... 400 mbar (nominal pressure) needs to be pro-
grammed.The liquid’s surface (with a density of 1) is 1 m above the diaphragm at the
zero point producing a pressure of 100 mbar.

This means that the device’s current value must be set to 8 mA when performing a dry
(empty) calibration.

5.4.2

Span Calibration

• Determine the hydrostatic pressure of the liquid’s surface, which corresponds to the

span end-point.

• Calculate the difference of the pressure value between span end-point and zero

point and divide this difference by the nominal pressure range of the sensor.

• Multiply this proportion by 16 mA and add 4 mA to the result.

This produces the calculated current (value I

calc

), which is entered in the transmitter

and used to program the span end-point (100%).
The measurement range between zero point and span end-point will be stored as
span.

I

calc

Zero point pressure (0%) 100 mbar

Sensors nominal pressure 400 mbar

------------------------------------------------------------------------------------------------

16 mA

4 mA

+

8 mA

=

=

Calibration Value

calc

Calibration of zero point, without display, without pressure

Calibration Value

... to I

meas

= I

calc

... to I

meas

= I

calc

(2 sec.)

(2 sec.)

(2 sec.)

026-DS-GB

This manual is related to the following products: