Overview of adjustments, Overview of adjustments -2 – Agilent Technologies E4438C User Manual
Page 94
Adjustments
Overview of Adjustments
4-2
Overview of Adjustments
In this section you will learn about all available adjustments for the Agilent E4438C ESG vector signal
generator. These adjustments are performed so that the signal generator is properly calibrated.
CAUTION
In all test equipment configurations, cables and adapters need to be properly torqued.
Exceeding recommended torque values may cause damage to a cable or adaptor and may cause
inaccurate test results.
Connector Type
Torque Value
APC 3.5
Use an Agilent 8710-1765 at 8-lb inch when
connecting APC 3.5 (m) to APC 3.5 (f).
SMA
Use an Agilent 8710-1765 at 8-lb inch when
connecting APC 3.5 (m) to SMA (f).
Use an Agilent 8710-1582 at 5-lb inch when
connecting SMA (m) to APC 3.5 (f).
Type-N
Type-N connectors may be connected finger tight. If a
torque wrench is used, Agilent 8710-1766 at 12-lb.
inch is recommended.