Fm scale dac offset calibration, Description, Required test equipment – Agilent Technologies E4438C User Manual
Page 107: Procedure, Fm scale dac offset calibration -15, Description -15, Required test equipment -15, Procedure -15
![background image](/manuals/42185/107/background.png)
Adjustments
FM Scale DAC Offset Calibration
4- 15
FM Scale DAC Offset Calibration
Description
This calibration is used to remove the offset associated with the FM SCALE DAC operational amplifier
located on the A17 Synthesizer (Option 501, 502, 503, 504). This calibration results in a DAC value for FM
OFFSET DAC 2. After this DAC value has been properly adjusted, the effects of the FM SCALE DAC value
on the offset will be minimized.
Required Test Equipment
• Agilent 53132A Option 050 frequency counter
Procedure
Figure 4-9
FM Scale DAC Offset Calibration Setup
1. Connect the equipment as shown above.
2. Preset all of the equipment.
3. Follow the instructions as they appear on the controller’s display.