Model HMP45C Temperature and Relative Humidity Probe

TABLE 4-2. Connections for Differential Measurements




Color




Description




CR10X

CR1000,
CR3000,

CR800,

CR5000,

CR23X


CR10,

CR510,

CR500




21X, CR7

Yellow Temperature

Signal

Differential

Input – H

Differential

Input – H

Differential

Input – H

Differential

Input – H

Jumper to

White

Temperature Signal

Reference

Differential

Input – L

Differential

Input – L

Differential

Input – L

Differential

Input – L

Blue

Relative Humidity Signal

Differential

Input – H

Differential

Input – H

Differential

Input – H

Differential

Input – H

White Signal

Reference

Differential

Input – L

Differential

Input – L

Differential

Input – L

Differential

Input – L

Black Power

Ground

G G G

Shield Shield G

G

Red Power

Continuous/Switched

12V/SW12V 12V/SW12V 12V/SW12V* 12V/SW12V*

CR10X Power Control

if using SW12V

channel on datalogger

Jumper from

SW12V

Control

to Control Port

*On these dataloggers switched power is only available with the SW12V peripheral.

TABLE 4-3. Power Connections using SW12V Peripherals

HMP45C SW12V

Peripheral

Datalogger

Color Description Terminal Wire

Red Power SW12V Red 12

V

Black Power

Ground GND

Black

G*

Green Control

Port


*The black wire of the SW12V should be connected to the type of datalogger
ground channel recommended for the HMP45C black wire as listed in Table
4-1 and Table 4-2.

5. Example Programs

This section is for users who write their own datalogger programs. A
datalogger program to measure this sensor can be created using Campbell
Scientific’s Short Cut Program Builder software. You do not need to read this
section to use Short Cut.