Appendix D. SDI-12 Sensor Support

7. To query the CS470/CS471 for its current SDI-12 address, enter the

command

?!. The CS470/CS471 will respond with the current SDI-12

address.

8. To change the SDI-12 address, enter the command

aAb!; where a is the

current address from the above step and b is the new address. The

CS470/CS471 will change its address and the datalogger will exit the SDI-

12 Transparent Mode.

9. Activate the SDI-12 Transparent Mode on Control Port 1 again by entering

1X . Verify the new SDI-12 address by entering the

?! command.

The CS470/CS471 will respond with the new address.

10. To exit the SDI-12 Transparent Mode, enter *.

FIGURE D-3. CR10X example of using the SDI-12 transparent mode to

change the SDI-12 address from 0 to 1. Sensor is connected to

control port 1.