CS470/CS471 Compact Bubbler System

FIGURE 7-3. Connection of the Bubble Tube

7. Install the NPT adapter on the end of the elbow or tee joint.

8. Activate the CS470/CS471 while immersing the bubble chamber. The

CS470/CS471 needs to be activated, so that the piston pump is operating

during this procedure.

9. Insert the tubing as far as possible into the bubble chamber’s tube fitting

10. Rotate the nut until finger tight.

11. While holding the bubble chamber steady, tighten the nut with a wrench

about one and one-quarter turns.

12. Install the bubble chamber on the NPT adapter.

For optimal measurement results, the bubble chamber must be

adjusted to be horizontal and aligned in the direction of flow

(maximum tolerance ±5°). The 25502 bubble chamber has a ball-

and-socket joint which allows for adjustments by 15° in any

direction.

7.6 Purging Function

On the front of the CS470/CS471, there is a pump membrane button (see

FIGURE 7-4). Pressing the button activates the purge function for as long as it

is pressed. The Status LED lights for approximately 2 seconds. With an

activated purge function, the sensor pumps a large amount of air through the