beautypg.com

D.2.3 cr10x datalogger example, D.2.3 – Campbell Scientific CS470/CS471 Compact Bubbler System User Manual

Page 41

background image

Appendix D. SDI-12 Sensor Support

In that case, simply enter the command SDI12 again, press ,

and key in the correct control port number when prompted.

6. To change the SDI-12 address, key in

aAb! , where a is the

current address from the above step and b is the new address (see

FIGURE D-2). The C470/CS471 will change its address and the

datalogger will respond with the new address. To exit SDI-12

transparent mode, select the

Close Terminal button.

FIGURE D-2. CR1000 example of using the SDI-12 transparent mode

to change the SDI-12 address from 3 to 1. Sensor is connected to

control port 1.

D.2.3 CR10X Datalogger Example

1. Connect a CS470/CS471 to the CR10(X) (see TABLE 7-2).

2. Download a datalogger program that contains the SDI-12 Recorder (P105)

instruction with valid entries for each parameter. Make sure that

parameter 3 of the P105 instruction matches the control port number where

the CS470/CS471 is connected.

3. In the LoggerNet Connect screen navigate to the

Datalogger menu and

select

Terminal Emulator. The terminal emulator window will open. In

the

Select Device menu, located in the lower left-hand side of the window,

select the

CR10X station.

4. Click on the

Open Terminal button.

5. Press the key until the datalogger responds with the

* prompt.

6. To activate the SDI-12 Transparent Mode on control port p, enter p

X

. For this example, key in

1X . The datalogger will

respond with entering SDI-12. If any invalid SDI-12 command is issued,

the datalogger will exit the SDI-12 Transparent Mode.

D-5