Ashcroft 2236 - Digital Sanitary Pressure Gauge User Manual

Page 30

background image

– 30 –

To replace the batteries (3˝ case):

1) Remove the single screw on the back of the gauge case.

2) Hold the keypad in the palm of hand.

3) Carefully remove the two batteries from the holder and

replace batteries. Use only AA alkaline non-rechargeable

batteries

To replace the batteries (4

1

2

˝ case):

1) Remove the ring on the front of the gauge case.

2) Looking at the gauge case, carefully pull the front face

out of the case.

3) Lay the gauge, face down on a flat surface.

4) Carefully remove the two batteries from the holder and

replace the batteries. Use only C alkaline non-recharge-

able batteries.

The Ashcroft digital sanitary gauge comes standard with a

Tri-Clamp connection. Install the Tri-Clamp fitting to the

mating fitting using a gasket compatible with the process

media and clamp (mating fitting, gasket, clamp provided

by others).

NEVER TIGHTEN GAUGE THREADS BY HOLDING THE

BODY OF THE GAUGE. DOING SO MAY DAMAGE THE

GAUGE AND MAKE THE GAUGE INOPERABLE.

Battery Installation and Replacement:

The Type 2032 comes standard with batteries installed.

The 3˝ case uses qty (2) AA alkaline batteries, the 4

1

2

˝ case

uses qty (2) C alkaline batteries. Use either Duracell

MN2400, MX2400 or Energizer E92BP, X92RP AAA alka-

line, non-rechargeable batteries.

Batteries have a life of approximately 1500 hours (3˝ case).

3600 hours (4

1

2

˝ case). Battery life is dependent on gauge

usage, backlite settings and power off settings. When the

lower bar of the battery icon of the gauge display flashes,

the gauge has approximately 7 hours of life remaining.

GAUGE INSTALLATION: