To use RECAL option:

Step 1: Press the Menu key on the keypad

Step 2: Press the

(up arrow key) or

(down arrow

key) on the keypad until the word CONFIG appears.

Step 3: Press Enter.

Step 4: Enter user password if it has been programmed.

Step 5: Press

(up arrow key) or

(down arrow key)

until the word RECAL appears.

Step 6: Press Enter.

Step 7: The gauge will now flash between INPUT and unit

of measure on the lower line and .00 on the top line. Apply

zero pressure to the gauge.

Step 8: Press Enter. Zero pressure is now set.

Step 9: The gauge will display full-scale pressure. Apply

full-scale pressure to the gauge.

Step 10: Press Enter. Full-scale pressure is now set.

Step 11: The gauge will now display mid-scale pressure.

Apply mid-scale pressure to the gauge.

Step 12: Press Enter. Mid-scale pressure is now set.

(Note: For compound ranges this recalibration is zero,

full-scale, mid-scale and full-vac.)

FOR FACTORY CALIBRATED SETINGS:

Step 13: To reinstate factory calibrated settings for zero,

full-scale and mid-scale press the

(down arrow key)

Menu key until the word FACT appears.

Step 14: Press Enter. Factory calibration settings are now

reinstated.

Step 15: After zero, full-scale and/or mid-scale or factory

default calibration have been set, the word SAVE appears

on the gauge display.

Step 16: Press Enter to finalize calibration.