Ashcroft 2236 - Digital Sanitary Pressure Gauge User Manual

Page 13

background image

Step 5: Press the ENTER key to finalize your choice. The

gauge display will now display SET. After two seconds the

screen will display the pressure value for 0% on the top

line. The middle line indicates the bar graph at 100% of

fullscale. The bottom line will display SET 0.

Step 6: Press the

(up arrow key) or

(down arrow

key) on the keypad to increase or decrease gauge value at

0% of range.

Step 7: Press the ENTER key to finalize your choice. The

new values for the bar graph and 4/20mA settings have

now been saved.

OFF: This option sets the amount of time before the gauge will

turn itself off. Offerings are Never*, 30MIN,10MIN, 5MIN, 2

MIN.

To use the OFF option:

Step 1: Press the MENU key.

Step 2: Press the

(up arrow key) or

(down arrow

key) until the word OFF appears.

Step 3: Press ENTER.

Step 4: Press the

(up arrow key) or

(down arrow

key) to select the desired OFF time.

Step 5: Press the Enter key to finalize the OFF setting.

UPDATE: This option allows for changing the rate at which

pressure is updated on the display screen. This feature is

useful with rapid changes in process pressure that may

cause flutter of the display. Options are 100ms*, 1 sec,

500ms and 200ms, updates per second or 100ms*.

Since customer processes vary, update rates should be

selected based on the application.

To use the UPDATE option:

Step 1: Press the MENU key.

– 13 –

MENU FUNCTIONS

*Indicates default.