AMETEK 7100 Leak Detect Stik User Manual
Page 6
![background image](/manuals/409750/6/background.png)
4
The type of information contained in each reading varies depending on the probe type. Table 2 specifies the data
pattern for various probe types.
Table 2 - Reading Data Type Specification by Probe Type
Types 1, 2, 3
Types 4, 5, 6
Reading 1
Temp 1
Temp 1
Reading 2
Product or Water*
Product or Water*
Reading 3
Low Ref Temp
Temp 2
Reading 4
Product
Product
Reading 5
High Ref Temp
Temp 3
Reading 6
Product
Product
Reading 7
Low Ref Temp
Temp 4
Reading 8
Product
Product
Reading 9
High Ref Temp
Temp 5
Reading 10
Product
Product
Reading 11
Circuit Temp
Circuit Temp
Reading 12
Product
Product
Reading 13
Low Ref Temp
Low Ref Temp
Reading 14
Product
Product
Reading 15
High Ref Temp
High Ref Temp
* For single float probes, this frame contains product data.
For dual float probes, this frame contains water data.
When calculating Water or Product positions, one must consider whether or not the probe uses a reference
magnet. For probes that use a reference magnet (type 3 and type 6), position can be calculated using the
following formula:
Position = Measured Period/Wire Speed
Where:
Measured Period = the time between pulses in a reading (ms)
Wire Speed = (ms / inch)
Position = distance from internal reference
For probes that do not use a reference magnet (types 1, 2, 4, and 5), position can be calculated using the
following formula:
Position = Measured Period/Wire Speed*2