beautypg.com

AMETEK 7100 Leak Detect Stik User Manual

Page 18

background image

16

Step 3 - Using the attached table, find the linear temperature interval (T

LIN1

, T

LIN2

) and then the actual

temperature interval (T

1

, T

2

) for T

LIN

.

Example.

T

LIN

= 15.763 °C,

T

LIN1

= 15.178 °C,

T

LIN2

= 16.321 °C,

T

1

= 14 °C,

T

2

= 15 °C.

Step 4 - Calculate accurate table interpolated temperature T, using linear interpolation within intervals (T

LIN1

,

T

LIN2

) and (T

1

, T

2

).

T = ((T

LIN2

- T

LIN1

) * T

2

- (T

LIN2

- T

LIN

) * (T

2

- T

1

)) / (T

LIN2

- T

LIN1

)

Example:

T

LIN

= 15.763 °C,

T

LIN1

= 15.178 °C,

T

LIN2

= 16.321 °C,

T

1

= 14 °C,

T

2

= 15 °C,

T = 14.512 °C