beautypg.com

Red Lion LD User Manual

Page 8

background image

8

COUNTER RESET AT POWER-UP





COUNTER B DECIMAL POINT POSITION

This selects the decimal point position for Counter B. The selection will also

affect Counter B scale factor calculations.

COUNTER B SCALE FACTOR

The number of input or batch counts is multiplied by the scale factor to

obtain the desired process value. A scale factor of 1.0000 will result in the
display of the actual number of input or batch counts. (Details on scaling
calculations are explained at the end of this section.)*



to











SCALING FOR COUNT INDICATION

The counter’s scale factor is factory set to 1, to provide one count on the

display for each pulse that is input to the unit. In many applications, there will
not be a one-to-one correspondence between input pulses and display units.
Therefore, it is necessary for the meter to scale or multiply the input pulses by
a scale factor to achieve the desired display units (feet, meters, gallons, etc.)

The Count Scale Factor Value can range from 00.0001 to 99.9999. It is

important to note that the precision of a counter application cannot be improved
by using a scale factor greater than one. To accomplish greater precision, more
pulse information must be generated per measuring unit. The following formula
is used to calculate the scale factor.

Scale Factor = Desired Display Units x Decimal Point Position

Number of Pulses

WHERE:
Desired Display Units
: Count display units acquired after pulses that occurred.

Number of Pulses: Number of pulses required to achieve the desired

display units.

Decimal Point Position:

0 =

1

0.0 =

10

0.00 = 100
0.000 = 1000
0.0000 = 10000



















USER INPUT ASSIGNMENT

The User Input Assignment is only active when Counter B is enabled and the

user input selection perfroms a Reset, Inhibit or Store function on one or both
of the counters.

USER INPUT FUNCTION















COUNTER A COUNT LOAD VALUE

Counter A resets to this value if Reset to Count Load action is selected. To

enter a negative Count Load value, increment digit 6 to display a “-” sign.*



to







* Indicates Edge Triggered function. All others are Level Active functions.

Advance once for each
activation.

Increase intensity one level
for each activation.

Reset setpoint 1 output.

Reset setpoint 2 output.

Reset both setpoint 1 and 2
outputs.

Freeze display for the selected
counter(s) while allowing
counts to accumulate internally.

Display Select *

Display Intensity Level *

Setpoint 1 Reset *

Setpoint 2 Reset *

Print and Reset *

Print Request

Setpoint 1 and 2 Reset *

See Programming Mode
Access chart (Module 3).

Edge triggered reset of the
selected counter(s) after
storing the count.

















Store and Reset



Store

Inhibit counting for the
selected counter(s).

Level active reset of the
selected counter(s).





Inhibit

Maintained Reset



Program Mode Lock-out

User Input disabled.

DESCRIPTION



No Function

DISPLAY

MODE

Serial transmit of the active
parameters selected in the Print
Options menu (Module 5).

Same as Print Request followed
by a momentary reset of the
selected counter(s).

EXAMPLE: The counter display is used to indicate the total number of feet

used in a process. It is necessary to know the number of pulses for the desired
units to be displayed. The decimal point is selected to show the resolution in
hundredths.

Scale Factor =

Desired Display Units x Decimal Point Position

Number of Pulses

Given that 128 pulses are equal to 1 foot, display total feet with a one-

hundredth resolution.

Scale Factor =

1.00 x 100

128
Scale Factor = 0.007812 x 100
Scale Factor = 0.7812

COUNTER B BATCH COUNT ENABLE





The Counter B Batch Count function internally counts the number of output

activations of the selected setpoint(s). The count source for the batch counter
can be SP1, SP2 or both. Batch counting is available in all count modes except
Dual Counter, which uses an external input signal for Counter B.

*For value entry instructions, refer to selection/value entry in the Programming

The Meter section.









The selected counter(s) will reset at each meter power-up.

Shaded area selections only apply when Counter B is enabled (Dual Count

mode or batch counter).