No lo-en, No fcs, 0 lo-t – Red Lion LDA User Manual

Page 8: 66 code, 50 code, 0 hi-t, No hi-en, Sec hi-en hi-t lo-en lo-t fcs code pro, Econdary, Unction

background image

8

Entering Code 66 will overwrite all user settings with

the factory settings. The meter will display

rESEt

and then

return to

CodE 00

. Press the

PAR

button to exit the module.

MIN DISPLAY ENABLE

YES

NO

NO

LO-En

NO

FCS

2.0

LO-t

MIN CAPTURE DELAY TIME

When the Input Display is below the present MIN value for the entered delay

time, the meter will capture that display value as the new MIN reading. A delay

time helps to avoid false captures of sudden short spikes.

0.0

to

999.9

seconds

Select

yES

to perform either of the Factory Service Operations shown below.

FACTORY SERVICE OPERATIONS

yES

NO

Enables the Minimum Display Capture capability.

The LD uses stored calibration values to provide accurate

measurements. Over time, the electrical characteristics of

the components inside the LD will slowly change with the

result that the stored calibration values no longer accurately

define the input circuit. For most applications, recalibration

every 1 to 2 years should be sufficient.

Calibration of the LD involves a calibration which should only be performed

by individuals experienced in calibrating electronic equipment. Allow 30 minute

warm up before performing any calibration related procedure. The following

procedures should be performed at an ambient temperature of 15 to 35 °C (59 to

95 °F).

CAUTION: The accuracy of the calibration equipment will directly affect the

accuracy of the LD.

Current Calibration

1. Connect the negative lead of a precision DC current source with an accuracy

of 0.01% or better to the COMM terminal. Leave the positive lead of the DC

current source unconnected.

2. With the display at

CodE 48

, press the

PAR

button. Unit will display

CAL NO

3. Press the

RST

button to select the range to be calibrated.

4. Press the

PAR

button. Display reads

0.0A

5. With the positive lead of the DC current source unconnected, press

PAR

.

Display reads

CALC

for about 8 seconds.

6. When the display reads the selected range, connect the positive lead of the DC

current source to the current input and apply full-scale input signal for the

range. (Note: For 200 mA range, apply 100 mA as indicated on the display.)

Press

PAR

. Display reads

CALC

for about 8 seconds.

7. Repeat steps 3 through 6 for each input range to be calibrated. When display

reads

CAL NO

, press the

PAR

button to exit calibration.

Voltage Calibration

1. Connect a precision DC voltage source with an accuracy of 0.01% or better

to the volt input and COMM terminals of the LD. Set the output of the voltage

source to zero.

2. With the display at

CodE 48

, press the

PAR

button. Unit will display

CAL NO

.

3. Press the

RST

button to select the range to be calibrated.

4. Press the

PAR

button. Display reads

0.0v

.

5. With the voltage source set to zero (or a dead short applied to the input), press

PAR

. Display reads

CALC

for about 8 seconds.

6. When the display reads the selected range, apply full-scale input signal for the

range. (Note: For 200V range, apply 100V as indicated on the display.) Press

PAR

. Display reads

CALC

for about 8 seconds.

7. Repeat steps 3 through 6 for each input range to be calibrated. When display

reads

CAL NO

, press the

PAR

button to exit calibration

CALIBRATION

66

CodE

RESTORE FACTORY DEFAULT SETTINGS

Entering Code 50 will display the model (LDA) and

version (x.x) of the meter. The display then returns to

CodE 00

. Press the

PAR

button to exit the module.

50

CodE

VIEW MODEL AND VERSION DISPLAY

48

CodE

5.2 Module 2 - s

econdary

F

unctIon

p

araMeters

(

2-SEC

)

PAR

Max Capture

Delay Time

Max Display

Enable

Min Display

Enable

Access Code

For Service

Operations

Min Capture

Delay TIme

Factory

Service

Operations

2-SEC

HI-En

HI-t

LO-En

LO-t

FCS

CodE

Pro

PARAMETER MENU

2.0

HI-t

MAX CAPTURE DELAY TIME

When the Input Display is above the present MAX value for the entered

delay time, the meter will capture that display value as the new MAX reading.

A delay time helps to avoid false captures of sudden short spikes.

0.0

to

999.9

seconds

NO

HI-En

MAX DISPLAY ENABLE

YES

NO

Enables the Maximum Display Capture capability.