Maintenance commands, Sta command, Table 21 – Vaisala FD12P User Manual
Page 85
Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 85
If the LM11 is not connected, the scaling factor should be negative.
Value -1.0 has been set at the factory as the default value. If a positive
value is used, the sensor expects a signal from the LM11.
For an example, see the following:
>blsc
BL SCALE
-1.000
>blsc 10.4
BL SCALE
10.400
If a day/night switch is connected to the serial line control input on the
FDP12 processor board, the FD12P can read the switch state and
report it as a background luminance value of 1 (day) or 0 (night). The
FD12P firmware will read the switch if the background luminance
scaling factor is set to 0.
Maintenance Commands
The maintenance commands are listed in Table 21 below.
Table 21
Maintenance Commands
Command
Description
STA
Displays status.
CAL
Calibrator_frequency Calibration.
TCAL
Temperature measurement calibration.
CLEAN
Sets clean references.
CHEC
Displays average signal.
FREQ
Displays internal signals.
DRY ON
Sets DRD12 dry offset.
WET ON
Sets DRD wet scale.
AN
CHANNEL
Analog channel (0,1,3,8 ... 15 or ANALOG ID).
STA Command
The STA command displays the results from the built-in test system
as a status message. Message 3 gives the same status message as the
STA command.
When you type
STA
↵
↵↵
↵
the system output is the following: