Message polling – Vaisala FD12P User Manual
Page 69

Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 69
Message Polling
In the polled (CLOSEd) mode, the FD12P sends a data message when
the host computer transmits a polling command. The message polling
mode is selected by the following command:
AMES Message_number Message_interval
↵
↵↵
↵
where
Message_number
- The valid range is 0 ... 7, refer to section Message Types on
- Selects the corresponding message as the default polled
message. Any negative message number is converted to 0.
Message_interval
- Negative or zero interval is used to disable the automatic
sending. This is used when messages are polled.
For example, AMES 0 0
↵ selects message 0, and cancels the
automatic sending.
The polling command format is the following:
where
=
ASCII character 05 hex
=
ASCII character 20 hex (space)
id
=
Selected in the configuration
message_
number
=
Optional
If only one FD12P unit in on the line and no id is set, the command
format is the following:
When the FD12P unit number two (id = 2) is polled for message
number 3, the command format is the following: