beautypg.com

Merit Medical Flex-Neck ExxTended Catheter User Manual

Page 9

background image

alternative measuring and cutting Instructions for the

upper chest catheter

1. position the upper catheter on the abdominal wall with the

black marker ring on the secondary incision, rectangle .

2. overlap the upper and lower catheters and choose an appro-

priate location to cut and join the two. See figure 16

note : keep in mind that one of the catheters will need to be

further shortened as described in step 5 below.

3. cut the two catheters at the location selected in step 2.

4. using the measuring rod, measure the distance from the

fascia to the skin at the primary incision site, at the skin surface

level.

5. cut either the upper or the

lower catheter (not both) to

further shorten it by the

distance measured in step 4.

note: In order to be able to

work with the lower catheter,

avoid using a length of

less than 5 cm.

Figure 16

connecting the lower and upper catheters

1. Wet the tapered tips of the double-barbed Titanium

connector in sterile saline or sterile water. do not use any other

lubricant.

2. Insert one end of the double-barbed Titanium connector

into the proximal end of the lower catheter. push the catheter

tip all the way to the connector’s center ring barrier. do not use

a twisting motion to force the catheter onto the connector. push

the connector into the catheter with a single forward motion.

do not dislodge the deep (rectus) cuff.

3. Insert the opposite end of the double-barbed Titanium con-

nector onto the ascending limb of the upper catheter, that is,

the limb with the marker ring, NoT the limb with the cuff. push

the catheter tip all the way to the connector’s center ring barrier.

do not use a twisting motion to force the catheter onto the

connector. push the connector into the catheter with a single

forward motion.

Note that for the upper chest exit-site, the lower and upper

catheters will be implanted with the same “side” up relative

to the positions of the radiopaque stripes. When inserting the

Titanium connector into the upper catheter, rotate the upper

catheter as necessary to ensure that the radiopaque stripes are

aligned.

4. check that radiopaque stripes are aligned. If the catheter

is implanted in the patient’s left side, both radiopaque stripes

will be face up (anteriorly). If the catheter is implanted in the

patient’s right side, both radiopaque stripes will be face down

(posteriorly).

5. Tie a non-absorbable suture, such as 2-0 or 0-polypropylene,

around each catheter, behind the barb on the Titanium connec-

tor. The two sutures may then be tied to each other to further

prevent tubing separation.

6. Test the integrity of the junctions by pulling firmly on each

catheter in turn while holding the connector. do not dislodge

the deep (rectus) cuff during this pull test .

Implanting the upper catheter with an upper chest exit-site

1. make a 2.0 - 3.0 cm horizontal incision at the marked second-

ary incision site, rectangle .

2. at the secondary incision site, rectangle , perform

dissection with a combination of hemostat clamps and ribbon

retractors to the pectoralis fascia. on the surface of the fascia,

create a subcutaneous pocket to contain the preformed arcuate

bend of the upper catheter. In addition, perform dissection

caudally in the retromammary space to facilitate passage of the

exxTended catheter Tunneling Tool tip from the primary (lower)

incision to the secondary (upper) incision.

3. Insert the blunt, bullet-shaped end of the Tunneling Tool into

the primary incision site.

4. Guide the Tunneling Tool along the surface of the fascia to

the secondary incision site.

note:

a. make sure that the Tunneling Tool stays in the relatively loose

avascular areolar tissue plane between the muscle fascia and

subcutaneous tissue. do not insert the tool and catheter into

the subcutaneous fat. doing so may cause the catheter to

kink during certain patient activities.

b. do not cross the patient’s midline of the abdomen or chest.

do not cross the sternum.

c. Follow the marked Tunnel Track, as indicated by the Stencil,

when feasible.

d. If a laparoscopic catheter implantation approach is being

used, the presence of a pneumoperitoneum provides a firm

fascial surface that facilitates passage of the Tunneling Tool.

5. advance the Tunneling Tool far enough through the second-

ary incision site so that it can be grasped with the other hand.

6. attach the proximal end of the upper catheter (the end

of the catheter closest to the cuff) to the barbed tip of the

exxTended catheter Tunneling Tool.

7. Secure the catheter end onto the tip with a suture.

8. carefully pull the Tunneling Tool out through the secondary

incision site far enough so that the Tunneling Tool can be laid

down.

note: do not twist the catheter. observe the radiopaque stripe

to ensure that the catheter remains straight.

9. continue to pull the catheter gently until the marker ring is

visible at the secondary incision site.

note:

a. When the marker ring is visualized on the surface of the fas-

cia at the secondary incision site, the length of the catheter

between the marker ring and the deep (rectus) cuff should

be relatively straight.

b. do not dislodge the deep (rectus) cuff.

c. do not twist or rotate the catheter. observe the radiopaque

stripe to ensure that the catheter remains straight. excess,

non-straightened tubing may cause future kinking and flow

failures under some conditions.

10. cut the catheter free of the Tunneling Tool.

note:

a. do not attempt to use the end of the catheter that was

inserted over the barbed tip of the Tunneling Tool. It is

stretched too much to be able to hold the connector

securely.

b. When cutting the catheter free of the Tunneling Tool, make a

straight, perpendicular cut of the tubing with suture scissors.

always verify that the cuts are perpendicular to the catheter

tubing so that the connector fits well in the catheter.

11. Infuse a minimum of 60 ml of sterile saline to verify patency,

and that there are no twists or kinks in the catheter.