beautypg.com

Merit Medical Flex-Neck ExxTended Catheter User Manual

Page 6

background image

2. record that distance, “a”, in centimeters, between the

primary and secondary incision sites. This distance, a, represents

the total length needed when combining upper and lower

catheters. See Figure 8.

3. Select and note a potential location where the upper and

lower catheters will be connected together by the internal dou-

ble-barbed Titanium connector. The Titanium connector should

rest on the flat surface of the abdominal wall fascia, avoiding

locations subject to pressure discomfort from the belt line and/

or angulation produced by flexing of the torso at skin creases.

note: When selecting the location to cut and join the cath-

eters, ensure that the connector will be positioned over the

abdominal wall, and not over the costal margin or chest.

4. using the provided

tape measure, measure

the distance between

the secondary incision

site, rectangle ,

and the chosen union

site. See Figure 9.

record that distance in

centimeters, as “B”. This

will be the length needed

of the upper catheter tubing

from the marker ring to the

catheter union site.

Figure 9

5. Subtract B from a to get the length of lower catheter tubing

needed to reach from the location of the deep cuff under the

fascia to the location of the catheter union. record that distance

in centimeters, as “c”. See Figure 9.

a - B = c , c is the total length needed for the lower catheter.

note: In order to be able to work with the lower catheter,

avoid using a length of c less than 5 cm.

6. Insert the measuring rod (located in the separate sterile

pouch along with the Tunneling Tool) beginning with the tip

marking 1 cm, into the primary incision next to the lower

catheter, perpendicular to the fascia and parallel to the catheter.

advance the measuring rod until the tip meets the rectus fascia.

See Figure 10.

Figure 10

7. hold the external limb of the lower catheter next to, and

parallel to, the measuring rod. using the centimeter markings

on the measuring rod, cut the lower catheter at the c length

above the level of the fascia (Figure 10). make a single, straight,

perpendicular cut of the tubing with suture scissors.

8. place the measuring rod alongside the upper catheter with

the B recorded length on the measuring rod at the level of the

catheter marker ring.

cut the upper catheter at the level of the tip of the measuring

rod Figure 11. make a single, straight, perpendicular cut of the

tubing with suture scissors.

Figure

11

alternative measuring and cutting Instructions for the

upper abdominal catheter

1. position the upper catheter on the abdominal wall with the

black marker ring on the secondary incision, rectangle .

2. overlap the upper and lower catheters and choose an appro-

priate location to cut and join the two. See figure 12.

note : keep in mind that one of the catheters will need to be

further shortened as described in step 5 below.

3. cut the two catheters at the location selected in step 2.

4. using the measuring rod, measure the distance from the

fascia to the skin at the primary incision site, at the skin surface

level.

5. cut either the upper or

the lower catheter (not both)

to further shorten it by the

distance measured in

step 4.

note: In order to be able

to work with the lower

catheter, avoid using a

length of less than 5 cm.

Figure 12

connecting the lower and upper catheters

1. Wet the tapered tips of the double-barbed Titanium con-

nector in sterile saline or sterile water. do not use any other

lubricant.

1. Insert one end of the double-barbed Titanium connector

into the proximal end of the lower catheter. push the catheter

tip all the way to the connector’s center ring barrier. do not use

a twisting motion to force the catheter onto the connector. push

the connector into the catheter with a single forward motion.

do not dislodge the deep (rectus) cuff.

2. Insert the opposite end of the double-barbed Titanium con-

nector onto the ascending limb of the upper catheter, that is,

the limb with the marker ring, NoT the limb with the cuff. push

the catheter tip all the way to the connector’s center ring barrier.

do not use a twisting motion to force the catheter onto the

connector. push the connector into the catheter with a single

forward motion.

note: When inserting the Titanium connector into the upper

catheter, rotate the upper catheter as necessary.

3. check that radiopaque stripes are 180° opposite. If the cath-

eter is implanted in the patient’s left side, the lower catheter

stripe will face up, while the upper catheter stripe will face

down. If the catheter is implanted in the patient’s right side, the

radiopaque stripes will be inverted: the lower catheter stripe

will face down, and the upper catheter stripe will face up. When