beautypg.com

55 series wideband usb sensor – Boonton The New 55 Series Wideband USB Power Sensor User Manual

Page 17

background image

55 Series Wideband USB Sensor

Page 17 of 28

Trigger/Time:

By selecting

icon you will be able to customize all related settings for both timebase and

trigger of a pulse signal.

Time base: Boonton software uses a fixed grid of 10 divisions
for the time-scale. The time resolution is set per division with
the lowest possible value of 5 ns/div.

Trigger Delay: Trigger delay can be adjusted both manually
entering numerical value into the field or using up-down
arrow keys.

Trigger Position: Trigger position can be changed by entering
numerical values into the field and also by dragging a
horizontal bar toward three default positions: Left, Middle
and Right.

Trigger sources and modes: Several trigger modes available
for each trigger source. Multiple trigger sources available
under the drop down list including both "Internal" and
"External" selection.

Trigger Mode: Four trigger modes available: Normal, Auto,
AutoLevel and Free run.

Trigger Level: Trigger level can be changed by using arrow
keys

Slope: Trigger slope can be selected by using positive and
negative options under slope.

Holdoff: Holdoff position allows user to select two different
modes: Normal and Gap.