beautypg.com

Lifestyle – AirSep MN112-1 User Manual

Page 94

background image

LifeStyle

AirSep

®

Corporation

• Adjustable shoulder strap
• Carrying bag

Before operating LifeStyle for the first time, allow the battery to charge for 12
hours. (Refer to the Battery Charging and Installation section.). Familiarise
yourself with the locations of the major components. These are illustrated in
the figures on the following pages and discussed later in this manual.

IV. 1.

Battery Charging and Installation

The battery must be charged before using LifeStyle. To charge the battery,
connect the AC adapter to the 12 VDC inlet on the side of the LifeStyle unit
(Figure 1) and into a 100-250 volt AC socket. Allow two and one-half hours for
charging the battery, unless the battery is being charged for the first time, then
the battery should remain in the unit for 12 hours with the unit connected into
an AC supply.

The figures below show how to install and remove the battery. Insert the
battery pack into the battery compartment with the arrow on top of the pack
pointing away from the LifeStyle unit’s handle. (See Figure 1.) To remove the
battery pack, push the pack in the direction of the arrow, and lift the battery out
of the battery compartment. (See Figure 2.)
• The battery pack must be in place in the unit when operating LifeStyle. If
LifeStyle will remain unused for more than five consecutive days, remove the
battery pack from the device, and store it in a cool, dry place.

IV. 2.

Nasal Cannula

A nasal cannula and tubing are used to
deliver oxygen from the LifeStyle unit to
the user. The tubing is connected to the
unit’s oxygen outlet. (See Figure 3.)

AirSep recommends a nasal cannula
with 2.1 m (7 ft) of tubing, AirSep Part
No. CU002-1.
• Use of some administration
accessories not specified for use with
this oxygen concentrator may impair
performance.

4-GB

Figure 1: Battery Insert

Figure 2: Battery Removal

12 VDC Inlet

Figure 3: Connecting Cannula

to LifeStyle’s Oxygen Outlet

GB