Lifestyle – AirSep MN112-1 User Manual

Page 106

background image

LifeStyle

AirSep

®

Corporation

16-GB

High and low pressure – audible and visual (RED alarm
light)

Sound:

Less than 55 decibels

Temperature range:

Operational temperature: 5°C to 40°C (41°F to 104°F)
Storage temperature: -20°C to 60°C (-4°F to 140°F) - unit
only
5°C to 35°C (41°F to 95°F) - battery only

*Based on an atmospheric pressure of 101 kPa (14.7 psia) at 21°C (70°F)

XV. 1. Classification

Type of protection against electric shock:

Class II Protection from electric shock is achieved by double

insulation.

Degree of protection against electric shock:

Type BF Equipment providing a particular degree of protection

against electric shock regarding
1) allowable leakage current;
2) reliability of protective earth connection (if present).
Not intended for direct cardiac application.

Independent testing for Medical Electrical Equipment Standard :

Tested by UltraTech Group of Labs to be in compliance with
IEC 60601-1:1988 + A1:1991, A2:1995

Protection against potential electromagnetic or other interference between the
equipment and other devices:

Tested by UltraTech Group of Labs to be in compliance with
IEC 60601-2 (EMC)

Method of cleaning and infection control allowed:

Please refer to the “Cleaning, Care, and Proper
Maintenance” section of this LifeStyle Patient Manual.

Degree of safety of application in the presence of flammable anesthetic
gases:

Equipment not suitable for such application.

Mode of operation:

Continuous duty

XV. 2. Symbols

Symbols are frequently used on equipment in preference to words with the
intention of decreasing the possibility of misunderstanding caused by
language differences. Symbols can also permit easier comprehension of a
concept within a restricted space.

GB