beautypg.com

Lifestyle, Gb xv. lifestyle specifications – AirSep MN112-1 User Manual

Page 105

background image

LifeStyle

AirSep

®

Corporation

Problem

Probable Cause Solution

Rapid alarm/pause/
alarm condition (three
½-second warnings
sound every 5
seconds), and the (red)
ALARM light illuminates
intermittently.

Rapid alarm condition
(½-second alarm
sounds every ½
second), and the (red)
ALARM light illuminates
intermittently.

Rapid alarm condition
(½-second alarm
sounds every ½
second), and the (red)
ALARM illuminates
continuously.

All other problems.

Breathing rate is
approaching the
threshold of the
LifeStyle unit’s capacity.

A general malfunction
could be imminent.

Breathing rate has
exceeded the capacity
of the LifeStyle unit.

A general malfunction
has occurred.

Reduce activity, and
consider locating another
source of oxygen.

Reduce activity, and
consider locating another
source of oxygen.

Change to another source
of oxygen, and contact your
Equipment Provider.

Change to another source
of oxygen, and contact your
Equipment Provider.

Change to another source
of oxygen, and contact your
Equipment Provider.

15-GB

XV. LifeStyle Specifications

Oxygen Concentration:*

Set at 1-5 lpm: equivalent to a continuous flow of
90% oxygen ± 3%

Dimensions:

13.9 cm high x 18.4 cm wide x 41.4 cm long (5.5 in.
high x 7.25 in. wide x 16.31 in. long)

Weight:

4.4 kg (9.75 lb)

Power:

LifeStyle unit – 13 VDC, 42 watts
Power supply – 100–250 VAC, 50/60 Hz, 2.0 amps max
Battery pack – 12 VDC, nominal

Battery duration:

50 minutes nominal for fully-charged battery

Battery quick-charge:

2-2 1/2 hours

Warm-up time:

3 minutes

Audible alarms and

Start-up – audible and visual (GREEN indicator light)

visual indicators:

Pulse flow – visual (GREEN pulse light)
Battery condition – audible and visual (YELLOW indicator,
warning, and alarm lights)
No breath detected – audible and visual (RED alarm light)
System overdraw – audible and visual (RED warning and
alarm lights)

GB