beautypg.com

Classification – AirSep MN170-1 D User Manual

Page 38

background image

MN170-1 rev D

34

Classification

Type of protection against electric shock:
Class II

Protection from electric shock is achieved by double insulation.


Degree of protection against electric shock:
Type BF

Equipment providing a particular degree of protection against electric shock regarding
1) allowable leakage current;
2) reliability of protective earth connection (if present).
Not intended for direct cardiac application.

Independent testing for Medical Electrical Equipment Standard:

Tested by QPS Testing Services NA Inc. to be in compliance with,

IEC 60601-1 Medical Electrical Equipment

– Part 1: General Requirements for Safety

Tested by QPS to be in compliance with applicable requirements of the Standard,

CAN/CSA C22.2 No. 60601-1-08 M90 Medical Electrical Equipment

– Part 1: General

Requirements for Safety

Protection against potential electromagnetic or other interference between the equipment and other devices.

Tested by Ultratech Group of Labs to be in compliance with:
EN 60601-1-2

Medical Electrical Equipment, Part 1: General Requirements for Safety-Collateral Standard:

Electrical Compatibility - Requirements and Tests
RTCA-DO160

Airborne Equipment, Sec. 21, Emission of Radio Frequency Energy

CISPR 11

/ EN 55011 Class B Group 1, “Industrial, Scientific, and Medical (ISM) Equipment”

FCC Part 15, Subpart B

– Class B Unintentional Radiators

Method of cleaning and infection control allowed:

Please refer to “Cleaning, Care, and Proper Maintenance” section of this Focus Patient Manual.


Degree of safety of application in the presence of flammable anesthetic gases:

Equipment not suited for such application.


Mode of operation:

Continuous duty.