How to Respond to Focus

’s Audible Alarm and Indicator Lights

Status

Audible Alarm

Light

Indicates

Action

Indicator


Brief, continuous at

start–up

(Green) and (Red)
alternate;

then (Green)

continuous light

Focus has been turned
on.

You may begin to operate your Focus
unit.

Indicator


No


(Green) flashes;

on each breath


Focus is delivering oxygen

as a pulse flow.


Continue using Focus normally.

Indicator


No


(Amber) continuous

light


Inspection and/or

Service

required.


Return unit to Equipment Provider

for inspection and/or

service.

Battery

Indicator


No


25% (Green) light;

flashes


Battery charge is low.


Connect the Focus unit into a DC

outlet or an AC outlet immediately.
Charge battery.

Alarm

Intermittent:
Beep

(Amber) alarm;
intermittent light

Warning: Battery voltage

approaching too low a
level to continue operating

Focus.

Connect the Focus unit into a DC
outlet or an AC outlet immediately.

Charge battery.

Alarm


Intermittent:
Beep, beep


(Amber) alarm;
intermittent light


Battery shutdown: Battery
voltage is too low to

operate Focus.


Connect the Focus unit into a DC
outlet or an AC outlet immediately.

Charge battery.

Alarm


Continuous:

Beep


(Amber) alarm;

continuous light


No breath detected by the

unit for a predetermined
time period.

Check the cannula connection. Ensure
that you are breathing through your

nose. If the alarm persists, contact
your Equipment Provider.

Alarm

Intermittent:
Beep, beep, beep

(Amber) alarm;
intermittent light

Breathing rate is
exceeding the capacity of

the Focus unit.

Reduce activity, and then if necessary
use another source of oxygen, as

available. Contact your Equipment
Provider.

Alarm

Audible alarm sounds
(beep,beep)

Audible alarm sounds
(beep,beep,beep)

(Red) light

General malfunction of the
Focus unit has occurred.

Turn off the unit. Change to another
source of oxygen, and contact your
Equipment Provider.

Check air inlet and exhaust vents for
obstruction.
Clear obstruction at air inlet and

exhaust.