beautypg.com

AirSep MN170-1 D User Manual

Page 34

background image

MN170-1 rev D

30

Focus Specifications

Oxygen Concentration:*

Pulse setting equivalent to a continuous flow of 90% oxygen - 3% / +5.5%

Pulse Dose:

17.25ml ± 10%

Dimensions:

6.4 in. high x 4.8 in. wide x 2.5 in. deep (16.4 cm high x 12.2 cm wide x 6.1 deep)

Weight:

Concentrator 1.75 lb (0.8 kg)

Battery 0.53 lb (0.2 kg)
Optional AirBelt Battery 1.8 lb (0.8 kg)


Power:

Universal Power Supply:
AC Power Supply: Input # 1_100 – 240 VAC (1.5 Amps max at 120 VAC 50/60 Hz)
DC Power Supply: Input # 2_11-16 VDC 5.0 Amps max

Battery duration
(Rechargeable
lithium Battery)

Battery: 1 ½ hours (per battery)
Optional AirBelt Battery: 4 hours


Battery recharge time:

4 hours; optional AirBelt: 3 hours

Warm-up time:

2 minutes


Battery cycle life:

Approximately 300 cycles, then 80% capacity or below.

Audible alarms and pulse
visual indicators:

Start-up – audible and visual
Pulse flow – visual
Cannula disconnect – audible and visual
Breath rate alarm – audible and visual
General malfunction – audible and visual
Service required – visual
Low battery – audible and visual

Battery condition – battery level indicator on battery

**Temperature range:

Normal operating temperature: 41°F to 104°F (5°C to 40°C)
Up to 95% RH (non-condensing)
Storage temperature:

-4°F to 140°F (-20°C to 60°C)

**Operational Altitude:

Operational Altitude: up to 10,000 ft. (3,048m) (523 mmHg) Higher altitudes may
affect performance



* Based on an atmospheric pressure of 14.7 psi (101 kPa) at 70°F (21°C)
**Operating outside of these operational specifications can limit the concentrator's ability to meet Oxygen Concentration specification

.

.