Agilent Technologies N3970A User Manual
Page 29
26
■
S
PECIFICATIONS
& O
RDERING
I
NFORMATION
Sensor Element
InGaAs
Wavelength Range
850-1650 nm
Calibration Wavelengths
850
1
, 1300, 1310, 1550 nm
Power Range
+5 to –60 dBm (850 nm)
+5 to –70 dBm (other wavelengths)
Maximum Input Power
+15 dBm
Uncertainty at reference conditions
2
±0.13 dB (±3%)
Total Uncertainty
3
±0.3 dB (±7%) ±2pW
Autotest sensitivity (typical)
4
-40 dBm (850 nm)
-50 dBm (other wavelengths)
Operating Time
250 hours
Applicable Fiber Type
9/125 µm to 62.5/125 µm
Optical Input
Non-Contact, Removable Connector Adaptors
Display Type
LCD, 4 digits
Display Resolution
0.01dB, 3 digits in linear mode
Display Units
dB, dBm, nW, µW, mW