Agilent Technologies N3970A User Manual
Page 25
22
■
C
ALIBRATION
& M
AINTENANCE
Enabling / disabling ‘shift’ keypad & re-booting the
microprocessor: See section on ‘getting started &
turning on.
Calibration
Calibration can be readily performed without opening the
instrument. Calibration is performed by setting up a singlemode
laser source at a stable but non-critical power level between -10
and -30 dBm, and adjusting the meter reading to the same value
as that shown by a reference meter.
A l t e rn a t i v e l y, known calibration constants can be
re-entered directly without using other equipment. This is useful
in case old calibration constants are to be put back, or the
instrument memory has been corrupted.
R e q u i red are laser light sources with accurate, calibrated
wavelengths and good power stability, a power meter with
appropriate calibrated wavelengths, singlemode test leads, an
anti-tamper label, and a 0.1” programming jumper. Check the
calibration certificates on your reference equipment, to ensure
current validity.
To enable calibration mode, open the battery compartment, hold
back the battery spring, remove the anti-tamper label, and insert
a 0.1” pitch programming shunt. Manipulation of the shunt is
easier with needle nose pliers. The meter will display ‘CAL’
Before commencing calibration:
■ Clean all connectors and the meter input ports very carefully
■ Ensure all devices have been at stable room temperature for
over an hour
■ The source is fully warmed up at the
λ
to be calibrated,
■ The meter battery is good.
On the meter to be calibrated, record the existing calibration
offsets as follows:
■ Select POWER METER, +/- to set the wavelength, and
MAX/MIN twice to display the offsets. Note them down.
(you can also re-enter or adjust known offsets at this point)