TOHO ELECTRONICS TTM-339 User Manual
Page 28

- 28 -
5.12
List of the alarm temperature setting mode setting
Name
Setting content
Initial value
1
E1H Event 1 upper limit setting
See Section 6.6 “Common parameter setting mode SET 5 to 8.”
2
E1L Event 1 lower limit setting
3
E2H Event 2 upper limit setting
4
E2L Event 2 lower limit setting
5
E3H Event 3 upper limit setting
6
E3L Event 3 lower limit setting
7
E4H Event 4 upper limit setting
8
E4L Event 4 lower limit setting
5.13
List of the PID setting mode setting
Name
Setting content
Initial value
1
P1 Proportional band setting for PID No. 1 (low temperature)
See Section 6.5 “Common parameter setting mode SET 4.”
2
I1 Integral time setting for PID No. 1 (low temperature)
3
D1 Derivative time setting for PID No. 1 (low temperature)
4
PM1 Intermediate point 1 setting for the PID range
5
P2 Proportional band setting for PID No. 2 (intermediate
temperature)
6
I2 Integral time setting for PID No. 2 (intermediate temperature)
7
D2 Derivative time setting for PID No. 2 (intermediate temperature)
8
PM2 Intermediate point 2 setting for the PID range
9
P3 Proportional band setting for PID No. 3 (high temperature)
10
I3 Integral time setting for PID No. 3 (high temperature)
11
D3 Derivative time setting for PID No. 3 (high temperature)
5.14
Caution
Care must be used for the names with suffixes as described below:
*1: Setting related to steps in a run cannot be changed.
*2: No change is possible during a run.
*3: SAVE (all settings writing) is performed.
6. Measurement range and indicator resolution
Input type
Standards
Measurement/measurement range
Indicator resolution
Thermocouple
K
JIS C 1602-1995
-200.0 to +1372.0
1
C / 0.1
C
J
JIS C 1602-1995
-200.0 to +1200.0
1
C / 0.1
C
T
JIS C 1602-1995
-200.0 to +400.0
1
C / 0.1
C
E
JIS C 1602-1995
-200.0 to +1000.0
1
C / 0.1
C
R
JIS C 1602-1995
-50 to +1768
1
C
S
JIS C 1602-1995
-50 to +1768
1
C
B
JIS C 1602-1995
0 to 1800
1
C
N
JIS C 1602-1995
-200.0 to +1300.0
1
C / 0.1
C
U
DIN
-200.0 to +400.0
1
C / 0.1
C
L
DIN
-200.0 to +900.0
1
C / 0.1
C
WRe5-26
ASTM
0 to 2300
1
C
PR40-20
ASTM
0 to 1880
1
C
PLII
ASTM
0.0 to 1390.0
1
C / 0.1
C
Resistance temperature
detector
Pt100
JIS C 1604-1997
-200.0 to +850.0
1
C / 0.1
C
J Pt100
JIS C 1604-1997
-200.0 to +510.0
1
C / 0.1
C
Voltage
0 - 1 VDC
-19999 to +29999
Display range of 20000
or less
Random change of
decimal point position
allowed
0 - 5 VDC
1 - 5 VDC
0 - 10 VDC
0 - 10 mVDC
Current
4 - 20 mADC