Specifications, Limited warranty – Setra System Model 470 User Manual
Page 34

34
Specifications
Pressure media:
Clean dry nonconductive noncorrosive gas
Accuracy:
accuracy
1
±
0.02% full scale at 70
°
F (21
°
C)
nonlinearity
±
0.012% full scale (End point)
hysteresis
0.010% full scale
nonrepeatability
±
0.010% full scale
thermal zero shift
2
±
0.002% FS/
°
F (0.004% FS/
°
C)
thermal span shift
2
±
0.001% FS/
°
F (0.002% FS/
°
C)
altitude resolution
1 ft. (4 ft. for 100 psi range)
stability, 24 hours
±
0.005% full scale
stability, 30 days
±
0.02% full scale
stability, 1 year
±
0.05% full scale
Output:
Bidirectional RS-232 interface
Physical:
pressure fitting
1/8” barbed male fitting
pressure port
10-32 internal thread
height
4.0”
width
3.5”
depth
5.25”
weight
2.4 pounds
Power:
5 VDC
±
1%, 70 mA; 17 mA w/200
microampere sleep mode option.
Notes:
1.
Accuracy as RSS of non-linearity, hysteresis, and non-repeatability.
2.
Unit calibrated at 70
°
F. Maximum thermal error is computed from this datum.
Information about the Model 470 Digital Pressure Transducer contained in this document is