Setra System Model 470 User Manual
Page 20

20
4.
Transmit S, for SETUP.
5.
If the elevation was in meters, transmit U, for CONVERT.
6.
Transmit S, for SETUP.
For example, if the station elevation is 1023 feet, transmit:
SB1023SS
Or, if the station elevation is 237 meters, transmit:
SB237SUS
7.
The Setra 470 is now in DASI mode and reports barometric pressure reduced to sea
level in response to an INTERROGATION command, indicated by “SEA LEVEL”.
8.
To switch from reporting true barometric pressure to corrected sea level pressure and
back, transmit B, for SEA LEVEL.
You must re-enter the station elevation (steps 1 through 6) if you move the transducer, or shut
off the power.
When you transmit B, for SEA LEVEL, in order to begin reporting pressure corrected to sea
level, the ZERO/TARE value (Section 4.8) is cleared (set to zero) automatically to ensure that
absolute sea level pressure is reported.
- Model ASM (8 pages)
- Model 204 (4 pages)
- Model 209 (4 pages)
- Model 227 (8 pages)
- 264 (4 pages)
- 269 (4 pages)
- Model 280E (4 pages)
- Model 370 (48 pages)
- Model 526 (7 pages)
- Model 540 (9 pages)
- Model 595 (8 pages)
- Model 869 (28 pages)
- Model 321 (4 pages)
- Model SRH (12 pages)
- Model Datum 2000 (31 pages)
- Model 217 (8 pages)
- Model 230 (12 pages)
- Model 242901-06 (1 page)
- Model 265 (4 pages)
- Model 270 (5 pages)
- Model 299 (4 pages)
- Model 524 (6 pages)
- Model 546 (8 pages)
- Model 730 (16 pages)
- GCT-225 (8 pages)
- Model 141 (4 pages)
- Model 205-2 (4 pages)
- Model 223 (8 pages)
- Model 231 (12 pages)
- Model 256 (8 pages)
- Model 276 (2 pages)
- Model 3100 (8 pages)
- Model 516 (6 pages)
- Model 550 (9 pages)
- Model 760 (18 pages)
- Model LD330 (4 pages)
- Model SRMD (8 pages)
- Model 201 (4 pages)
- Setra Model 206 (4 pages)
- Model 224 (8 pages)
- Model 231RS (16 pages)
- 260 (2 pages)
- 267MR (14 pages)
- Model 278 (2 pages)