beautypg.com

8 technical data, 1 specifications, Technical data – ROTRONIC HygroLab C1 User Manual

Page 25: Specifications, E-m-hylabc1-v1, 8technical data, Hygrolab c1 bench-top indicator: user guide

background image

E-M-HyLabC1-V1

Rotronic AG
Bassersdorf, Switzerland

Document code

Unit

HygroLab C1 bench-top indicator:

User Guide

Instruction Manual

Document Type

Page

25 of 28

Document title

© 2011; Rotronic AG E-M-HyLabC1-V1

8

Technical data

8.1 Specifications

General

HygroLab C1

Device type

Humidity-temperature bench-top indicator with 4 probe inputs
(HygroClip 2 digital probe) and real time clock (date and time)

Power

Model AC1211 power adapter (100…240 VAC – 12 VDC, 200 mA).

The USB interface can be used to power the HygroLab C1 (no power
adapter required - current draw: about 70 mA). Use of the Ethernet
interface always requires a power adapter.


Operating modes

HygroLab C1

AW measurement

Display of water activity and temperature

o

AwE

: conventional measurement

o

AwQuick : accelerated measurement

Standard mode

Display of %RH, temperature + calculated parameter
Display of analog probe measurement


Probe inputs

HygroLab C1

Number of probe inputs

4

Probe type

HygroClip 2 digital humidity temperature probe


Humidity and temperature
measurement

HygroClip 2 probe

See document E-M-HC2 Probes > Specifications


Calculated parameters

HygroLab C1

Psychrometric calculations

Dew point (Dp) above and below freezing
Frost point (Fp) below freezing and dew point above freezing
Wet bulb temperature (Tw)
Enthalpy (H)
Vapor concentration (Dv)
Specific humidity (Q)
Mixing ratio by weight (R)
Vapor concentration at saturation (Dvs)
Vapor partial pressure (E)
Vapor saturation pressure (Ew)


Start-up time and data refresh rate

HygroLab C1

Start-up time

3 s (typical)

Data refresh rate

1 s (typical)


Functions

HygroLab C1

Water activity measurement

o

Conventional water activity measurement

o

Accelerated water activity measurement

o

Automatic end of measurement

o

Simultaneous measurement of up to 4 probes