beautypg.com

E-m-hylabc1-v1, Hygrolab c1 bench-top indicator: user guide – ROTRONIC HygroLab C1 User Manual

Page 16

background image

E-M-HyLabC1-V1

Rotronic AG
Bassersdorf, Switzerland

Document code

Unit

HygroLab C1 bench-top indicator:

User Guide

Instruction Manual

Document Type

Page

16 of 28

Document title

© 2011; Rotronic AG E-M-HyLabC1-V1

4. AWQ Ended: when the projection is stable for a probe being displayed, the HygroLab C1 automatically
ends the measurement and freezes the display for that probe. A check mark appears to the left of the water
activity value. In addition the HygroLab C1 can be set with the HW4 software to emit an acoustic signal (Beep)
lasting 5 seconds when the last probe is done. The acoustic signal can be stopped by pushing a key (do not
press the ENTER key).

When several probes are connected to the HygroLab C1, use the UP or DOWN key to view each probe.

IMPORTANT: do not press the ENTER key until you have reviewed each probe.



5. AWQ Reset: write down the measurement for each probe or use the automatic data capture function. This
function can be enabled from the keypad: MENU > Aw Settings > SaveResult > ON. The Aw value measured
by probe 1 is saved to Data Bin 1. Probe 2 is saved to Data Bin 2, etc.

Press ENTER. The HygroLab C1 is ready to start a new measurement

Measurement ended

Final
Aw projection