beautypg.com

E-m-hylabc1-v1, Hygrolab c1 bench-top indicator: user guide – ROTRONIC HygroLab C1 User Manual

Page 10

background image

E-M-HyLabC1-V1

Rotronic AG
Bassersdorf, Switzerland

Document code

Unit

HygroLab C1 bench-top indicator:

User Guide

Instruction Manual

Document Type

Page

10 of 28

Document title

© 2011; Rotronic AG E-M-HyLabC1-V1

Main Menu

Menu Items

Submenu
Items

Selections or
Information

Notes

Probe Settings

Calc 1… 4

Psychrometric
Parameter. See 2.5

Calculated parameter
Probe 1, 2, 3, 4

Formula 1…4

ON / OFF

User defined calculation
Probe 1, 2, 3, 4
(requires HW4 software)

When ON, this calculation
replaces the psychrometric
parameter both on the
local on the display and on
a PC monitor (HW4
software)

Probe Actions
(select probe
input first)

Probe Info

Serial Number
Version
Probe Name

Humi Adjust

Acquired Points

Lists the cal. points
present in the probe
memory (a max. of
10 points are
shown)

Additional Options:

o

Clear all cal. points

o

Clear the last point

Acquire Ref
Probe

Saves value measured by
the reference probe as a
cal. point

Acquire Ref
Value

Manual entry:
known reference
environment

Saves value entered
manually as a cal. point

Start
Adjustment

Adjusts the probe

Effect depends on number
of calibration points in
probe memory

Reset to
Factory

Returns the probe to the
initial factory adjustment

Temp Adjust

Acquired Points

Lists the cal. points
present in the probe
memory (a max. of 2
points are shown)

Additional Options:

o

Clear all cal. points

o

Clear the last point

Acquire Ref
Probe

Saves value measured by
the reference probe as a
cal. point

Acquire Ref
Value

Manual entry:
known reference
environment

Saves value entered
manually as a cal. point

Start
Adjustment

Adjusts the probe

Effect depends on number
of calibration points in
probe memory

Reset to
Factory

Returns the probe to the
original factory adjustment

Set as
reference Probe

Sets the selected probe as
the reference probe