4.5 Using the AwE mode
The HygroLab C1 monitors the water activity and temperature values measured by the probe. When both
values are at equilibrium during a few minutes, the measurement is automatically ended. Depending both on
the product being measured and on the stability of temperature, measurements typically require 30 to 60
minutes.
1. AWE Reset: the HygroLab C1 is ready to start a measurement using up to 4 probes
When ready to measure, press on the ENTER key.
2. AWE Running: the HygroLab C1 starts monitoring the humidity and temperature signals for equilibrium
Current
humidity value
Current
temperature value
AwE mode is running
Probe 1
is being displayed
Elapsed time
hh:mm:ss