3. AWE Ended: when both the humidity and temperature signals measured by a probe are stable, the
HygroLab C1 automatically ends the measurement and freezes the display for that probe. A check mark
appears to the left of the water activity value. In addition the HygroLab C1 can be set with the HW4 software
to emit an acoustic signal (Beep) every 5 seconds when the last probe is done. The acoustic signal can be
stopped by pushing a key (do not press the ENTER key).
When several probes are connected to the HygroLab C1, use the UP or DOWN key to view each probe.
IMPORTANT: do not press the ENTER key until you have reviewed each probe.
4. AWE Reset: write down the measurement for each probe or use the automatic data capture function. This
function can be enabled from the keypad: MENU > Aw Settings > SaveResult > ON. The Aw value measured
by probe 1 is saved to Data Bin 1. Probe 2 is saved to Data Bin 2, etc.
Press ENTER. The HygroLab C1 is ready to start a new measurement
Measurement ended
Final
Aw measurement