Performing a flash update
Whenever an update is available, a flash update of the docking station internal Ethernet module
can be done as explained below:
• Be sure to have the IP address of the docking station available (example: 192.168.5.76)
• Place a HygroLog NT on the docking station and keep the docking station powered during
the entire flash update process
• Download a copy the update file (example: dsu4web_vxx.bin) to the root directory of your
PC hard drive (C:\)
• In Windows XP, click on START > All Programs > Accessories > Command Prompt
• In Command Prompt, type CD \ (changes the directory to the root of drive C)
• At the command prompt C:\ type the following sequence of commands as indicated in bold
characters below (any text in italics denotes a PC answer):
C:\>ftp 192.168.5.76
Connected to 192.168.5.76
220 NET+OS 6.3 FTP server ready
User (192.168.5.76:(none)): (none) - note: include the brackets
230 User (none) logged in
ftp> bin
200 Type set to I.
ftp> put dsu4web_vxx.bin (use the actual bin file name)
200 PORT command Ok.
150 About to open data connection.
226 Transfer complete (this answer may take some time to appear)
FTP: 385940 Bytes sent in 1.58Seconds 244.58 Kbytes/sec.
ftp> quit
221 Goodbye.
• C:\>exit (closes Command Prompt and return to Windows)