8.2 Additional information

• The Web Server graph data is lost when power is turned off.
• When the IP address of the data logger is accesses via Internet, be cautious when

synchronizing the data logger time and date with your PC since you may not be in the same
time zone as the data logger.

• After synchronizing the data logger time and date with the PC, go to Web Server Settings and

click on “Apply”. This erases all data from the docking station volatile memory and prevents
potential conflict between the old and new time.

• Except for the log file header, HW4 is required to read the data from any downloaded log file

in binary (LOG) format.

• HW4 is required to view the data from any downloaded log file in graphic form.
• The data logger cannot be interrogated or written to during a file download. Measurement

data displayed by the Web Server is not updated.

• Direct connection via TCP Port 2101: to maintain compatibility with HW4, connectivity with

the data logger is possible using the TCP Protocol. This function cannot be disabled or
password protected.

• RS-485 network: not supported, with or without HW4.