In-e-ds-u4web-v1_14 – ROTRONIC Dockingstation with WEB-Server User Manual
Page 11
IN-E-DS-U4WEB-V1_14
Rotronic AG
Bassersdorf, Switzerland
Document code
Unit
Docking station DS-U4WEB with web
server version 1: instruction manual
Instruction Manual
Document Type
Page
11 of 15
Document title
© 2007; Rotronic AG
IN-E-DS-U4WEB-V1_14
HTML
Description
Home.htm
(Current Values)
Rotronic Logo, firmware version, page navigator and current
measurement data for the inputs selected in the Web Server Settings
page. Automatic refresh rate: 60 seconds.
Messwerte.htm
This page is similar to Home.htm without Rotronic logo, firmware
version and page navigator. Automatic refresh rate: 60 seconds.
Probes.htm
Displays the current measurement data for all inputs (selected or not)
as unformatted text. (HTML)
Humi1 ... Humi7
(no extension)
Current humidity value for specified input (incl. unit) in text format
Temp1 … Temp7
(no extension)
Current temperature value for specified input (incl. unit) in text format
Calc1 … Calc7
(no extension)
Current calculated parameter value for specified input (incl. unit) in text
format
MesswerteChart.htm
Displays a bar graph for the parameter selected in the page header
table. The most recent 500 values (at the time the page was opened)
are displayed in the graph, based on the graph interval selected under
Settings. Bring the mouse cursor over the chart to see a data cursor
that displays the value, date and time of one data point at a time.
- Automatic scale
- No zoom function
- No automatic refresh
Note: the data underlying the chart is held in the docking station
volatile memory as opposed to the flash memory card of the data
logger. This data is lost when power is removed from the docking
station or when clicking on the apply button in the Web Server Settings
page.
Properties.htm
Displays the current configuration settings for both the data logger and
docking station. The settings can be changed from this page.
FreeFlash.htm
Displays the amount of free memory available on the data logger flash
card
Recording.htm
Displays the current settings of the data logger recording function. The
settings can be changed from this page