Section 1: Device Safety

CAUTION: DEVICE MUST BE USED ONLY IN A MANNER CONSISTENT
WITH THIS MANUAL.

NOTICE: WHEN 230VAC SUPPLY USED, DWYER SPECIFIES USE OF
AC SUPPLY PART # T35-9-100R-3 MANUFACTURED BY ENG
ELECTRIC. THIS SUPPLY IS AVAILABLE FROM DWYER.

1.1 Model information

• Model:

DLI-R

Description: Single channel RTD logger with LCD display

Manufactured in the USA

1.2 Specifi cations

External Probe

Range: -328 to 1562ºF (-200 to +850ºC)

Resolution:

0.01ºF

(0.01ºC)

Calibrated accuracy (case between 59 to 95ºF (15 to 35ºC))

±0.05ºF (±0.05ºC) Probe between -328 to 500ºF (-200 to 260ºC)

±0.1ºF (±0.1ºC) Probe between 500 to 1562ºF (260 to 850ºC)

Accuracy de-rating:

0.05ºF (0.05ºC) per ºF (ºC) outside 59 to 95ºF (15 to 35ºC ) case

temperature, probe between -328 to 500 (-200 to 260ºC)

0.1ºF (0.1ºC) per ºF (ºC) outside 59 to 95ºF (15 to 35ºC) case

temperature probe between 500 to 1562ºF (260 to 850ºC)

Dot-Matrix LCD

Dimensions: 2.5” x 1.375” (63mm x 35mm)

Text:

Confi gurable channel text size

Indicators: Power, status, memory

Backlight:

Confi gurable w/auto shut-off and contrast adjustment

Start/Stop Time:

Software programmable start time and date, up to six months in

advance; programmable stop time

Start/Stop Time:

Software programmable start time and date, up to six months in
advance; programmable stop time