beautypg.com

Input ranges (field selectable), Thermocouple types, Rtd types – Dwyer 16L User Manual

Page 26: Process input types

background image

949-1270

Page 26 of 28

December, 1998

Input Ranges (Field Selectable)

Thermocouple Types

Input
Type

Range

1°F

1°C

Type J or L*

-100 to +1600

-73 to +871

Type K*

-200 to +2500
-129 to +1371

Type T*

-350 to +750
-212 to +398

Type E*

-100 to +1800

-73 to +982

Input
Type

Range

1°F

1°C

Type R

0 to 3200

-17 to +1760

Type S

0 to 3200

-17 to +1760

Type B

+75 to 3308
+24 to 1820

Type C

0 to 4208

-17 to 2320

Input
Type

Range

1°F

1°C

Type N*

-100 to +2372

-73 to +1300

RTD Types

Input
Type

Range

1°F

1°C

100 Ohm

Platinum

0.00385 DIN

Curve*

-328 to +1607

-200 to +875

100 Ohm

Platinum

0.00392 Nist

Curve*

-328 to +1607

-200 to +875

120 Ohm

Nickel

0.00628 US

Ind. Curve*

-112 to +608

-80 to +320

1000 Ohm

Platinum

0.00385 Nist

Curve*

-328 to +1607

-200 to +875

Process Input Types

* These Input Types can be set for 0.1° display.
If temperature goes above 999.9° or less than
-199.9° the display will return to whole degree
resolution.

The 0 to 20 mAdc, 4 to 20 mAdc, 0 to 10 Vdc, 2 to 10 Vdc, and -10 to +10
mVdc inputs are fully scalable from a minimum of 100 counts span placed
anywhere within the within the range of -1999 to +9999. Decimal point
position is adjustable from the zero place (9999), tenths (999.9), hun-
dredths (99.99), thousandths (9.999), or ten thousandths (.9999).