Functions
3
XP2i Operation Manual
Functions
Units
Button
Pressing this button causes the WT Series to select the next available unit of pressure measurement.
See
Pressure Ranges, Display Scales, & Resolution on page 14
or the list of pressure units available for your model.
Units that you don’t need or never use can be turned off . You can also defi ne a special unit for your WT Series with Confi gXP.
You can use the WT Series to display directly in a unit not otherwise available, such as feet of seawater, or foot-pounds of torque. When you select your custom
unit from the keypad, the screen displays the USER icon.
Peak
Button
Pressing the
(
peak
)
button causes the display to cycle through the following , depending on your setting in Confi gXP.
. . . . Maximum detected pressure
. . . . Minimum detected pressure
(blinking)
. . . . PSV Mode, maximum
*
(blinking)
. . . . PSV Mode, minimum
*
. . . . Average pressure
. . . . DataLoggerXP datalogging mode
*
From the factory this setting is disabled. Use Confi gXP to enable.
Peak High and Peak Low values are not saved when the gauge shuts off ; they will reset to the current reading when the WT Series is turned on or reset.
In some cases the ability to display a peak value may not be needed, or may even be dangerous. Confi gXP allows you to disable this button.
The WT Series can average 1 to 10 readings, recalculated every time pressure is measured (4 times per second). Enable and set the number of readings to be averaged with Confi gXP.