M2000 Harness Connections
1. Clip the alligator clip labeled Detector Ground to either of the hex nuts mounted on top of the meter flanges.
2. Plug the connector labeled Amplifier Electrode into the circuit board connector labeled E1, ES, E2, RS, EP, ES.
3. Plug the Amplifier Coil Output into the circuit board connector labeled CS, C2, C1.
4. Plug the Output 1 & 2/ RS232 into the board output connector labeled 1 to 7.
5. Plug the Output 3 & 4 / Input into the board output connector labeled 1 to 14.
6. Plug the Analog Output connector into the board connector labeled 15 and 16 in connector row of
COMMUNICATION / Analog Out on the right side.
7. Connect the harness wire connector labeled Detector Electrode to the 6-wire connector from the detector.
8. Connect the harness wire connector labeled Detector Coil to the 3-wire connector from the detector.
BA
20
00-
22
M
ID
L
N
PE
JP1
JP2
COMMUNICATION
DISPLAY
CS C2 C1
E1
ES
E2
RS
EP
ES
15 16
1
2
3
4
5
6
7
8
9
10
11
12
13
14
Output 1 & 2 / RS232
Output 3 & 4 / Input
Analog Output
Amplifier Coil
Amplifier Electrode
Figure 5: M2000 Wire Harness Circuit Board Connectors
Detector Electrode
Detector Coil
Detector Ground