Field verification device functions, Verification device kit components – Badger Meter M-Series Field Verification Device User Manual
Page 6
Page 6
8-11
Field Verification Device
Field Verification Device Functions
The Device also:
•
Determines if the M-Series meter is within one percent of the original factory calibration.
•
Verifies the functionality of all of the meter's inputs and outputs.
•
Measures electrode resistance and integrity.
•
Measures coil resistance and integrity.
•
Measures coil insulation resistance.
•
Measures current and frequency at previously selected flow rate.
•
Evaluates the signal processing functionality.
•
Verifies that there is no high-voltage noise disturbing the coil circuit.
Verification Device Kit Components
The Field Verification Device is packaged in a foam-lined, durable plastic case and includes the following components.
1. One (1) Verification Device +5 VAC, 3.0A power adapter
2. Four (4) AC power conversion connectors
3. One (1) USB PC data cable
4. One (1) DC power adapter
5. Two (2) Verification cable harnesses: one for the M2000 and one for the M5000
Figure 1: Kit Components
1
2
3
4
5