West Control Solutions KS 800 ISO1745 Interface Manual User Manual
Page 15
Bem. A Unit_State1
MSB
LSB
D7
D6
D5
D4
D3
D2
D1
D0
Bit no.
Name
Allocation
Status ‘0’
Status ‘1’
D0
‘0’
Always ‘0’
D1
CNF
Instrument status
on-line
configuration
D2...D4 ‘0’
Always ‘0’
D5
UPD
Parameter update
no
yes
D6
‘1’
Always ‘1’
D7
Parity
Bem. B DPErr
MSB
LSB
D15 D14 D13 D12 D11 D10
D9
D8
D7
D6
D5
D4
D3
D2
D1
D0
Bit no.
Name
Allocation
Status ‘0’
Status ‘1’
D0
Buszugriff nicht erfolgreich
Kein Fehler
Fehler
D1
Fehlerhaftes Prarametriertelegramm
Kein Fehler
Fehler
D2
Fehlerhafte Konfiguration
Kein Fehler
Fehler
D3
Kein Nutzdatenvberkehr mer
Kein Fehler
Fehler
D4...D15 ‘0’
Always ‘0’
Bem. C HWbas
COM2
0
0
T
H
Z
E
Basic version without COM2
0
0
0
0
COM2 with CANopen
0
1
0
0
COM2 with PROFIBUS-DP
1
0
0
0
COM2 with ISO1745
1
1
0
0
Example: Value ‘HWbas = 0100‘ means that the addressed instrument is a COM2 interface with CANopen
connection.
Bem. D SWopt
Version
0
0
T
H
Z
E
Basic version
0
0
0
0
Water cooling (so far not available)
0
1
0
0
Bem. E SWCod
T
H
Z
E
7th digit
8th digit
9the digit
10th digit
Example: Value ‘SWCod= 7239‘ means that the software for the addressed instrument contains code
number 4012 157 239xx.
Bem. F SWvers
T
H
Z
E
0
0
11th digit
12th digit
Example: Value ‘SWVers= 11‘ means that the software of the addressed unit contains code number 4012
15x xxx11.
Bem. G UPD
Changing a parameter value or a configuration value via an interface is indicated in the UPD flag. This bit is
also set after mains recovery. The flag, which can be read also via code UPD, can be reset (value =0).
Function block protocol
15