beautypg.com

Thermo Fisher Scientific CyberScan pH 1000/pH 2500 User Manual

Page 9

background image

Instruction Manual

pH 1000/ 2500

4

CAL/MEAS

Primary Function

Press to toggle between measurement and calibration modes of the meter - pH,
Temperature, mV, Relative mV and Ion. Also use as exit from SETUP mode,
secondary functions and to abort operation while in calibration mode.

Secondary Function

None

Numeric Value

None

1 / MODE

Primary Function

Select one of four measurement modes: pH/mV/Relative mV/Ion.

Secondary Function

None

Numeric Value

1

2 / HOLD

Primary Function

Freeze the displayed value. Allow you to print the held reading or store it into memory.

Secondary Function

None

Numeric Value

2

3 / SETUP / REL MV BASE

Primary Function

Enter SETUP mode of the meter for customization of meter functions as well as view
some diagnostic functions.

Secondary Function

Display the Relative mV base value while in Relative mV mode and Ion Electrode
Slope while in the Ion mode.

Numeric Value

3

4 /

/ RESLN

Primary Function

Select next higher pH buffer values in the pH calibration mode. Also used to scroll
through SETUP mode options.

Secondary Function

Toggle the pH resolution between 0.01 pH and 0.1 pH.

Numeric Value

4

5 /

/ SLOPE

Primary Function

Select the next lower pH buffer value option in pH calibration mode. Also used to
scroll through some SETUP mode options.

Secondary Function

Display the percentage slope of the pH electrode when in the pH mode. While in Ion
mode, it displays slope of ion electrode in mV/decade.

Numeric Value

5

6 / SET TEMP / OFFSET

Primary Function

Press to calibrate ATC temperature probe or to set manual temperature.

Secondary Function

Displays the offset of pH electrode. This is useful as a diagnostic function to
determine the quality of the probe that is connected to the meter.

Numeric Value

6