Thermo Fisher Scientific CyberScan pH 1000/pH 2500 User Manual

Page 29

background image

Instruction Manual

pH 1000/ 2500

23



12.7 Setting the High and Low Setpoint Alarm

(pH/mV/Relative mV)

An alarm can be set to give an audible warning to you whenever
the set limits for pH, Relative mV or mV has been exceeded.

Hint:

You can set individual high and low setpoints

for each of the measurement modes of pH,
Relative mV and mV. The individual setpoint alarms
will be automatically activated whenever the
measurement mode of the set alarm is entered.

1. Press

ALARM. The meter will prompt for the setting or changing of

the setpoint. If no setpoint is set, the meter displays “- - - -”.
Otherwise the old setting is shown.

2. Enter the high setpoint using the numeric keys. To erase a setpoint,

press +/- key twice.

3. Confirm

using

YES key. To cancel and re-enter press NO. Meter will

advance to low setpoint upon confirmation of value.

4. Repeat steps 2 and 3 for the low setpoint setting. Meter then reverts

to the measurement mode.

Hint

: To shut off an alarm, go back into the ALARM

function and press +/- key twice at the setpoint
options. The alarm will then shut off.






12.8 Setting & Activating the Timer

The meter can be set to hold a reading at a particular time until it
is read. This is done by first setting the timer countdown and
activating it using the ACTIVATE key. At the time set, the meter
holds the reading (until any key is pressed and gives a continuous
beep to signal to you that the time set is reached.

If you wish to skip input for

either the High/Low

setpoint, press YES key.

Meter automatically exits to

measurement mode.