beautypg.com

5turning the meter on and off – Thermo Fisher Scientific CyberScan pH 1000/pH 2500 User Manual

Page 12

background image

Instruction Manual

pH 1000/ 2500

7

5

TURNING THE METER ON AND OFF


Once the AC/DC adapter is connected and powered on,
the meter beeps and the full LCD lights up for a few
seconds to display all segments as a self-diagnostic
test of the LCD. Then the meter will display the current
time.
To access the meter functions, press ON/OFF key. The
meter beeps and performs a diagnostic test before
going into the pH measurement [MEAS] mode. To turn
the meter off, press ON/OFF key within the operation
mode. The meter performs a self-diagnostic LCD test
before displaying the current time.

IMPORTANT :

Remember to remove the protective cap from
the electrode tip prior to measurement.
Replace it in place after use or store electrode
in some storage solution to avoid possible
dehydration of glass bulb which may result in
slow response.

In any case if the electrode is not in use for
long period, condition the electrode by
soaking into tap water for 1-2 hours before use.

Make sure to wash electrodes/probes
before/after each sample measurements to
prevent any carry-over of samples.