Thermo Fisher Scientific CyberScan pH 300/310 User Manual

Page 49

background image

46

13 ADDENDUM 1: METER FACTORY DEFAULT

S E T T I N GS

Resetting the meter to factory default settings clears all calibration data and most other setup
functions you might have changed. The following settings will remain as you have set them:

Temperature unit of measure (°C or °F) – only for pH 310 meter

The temperature offset calibration value.

• Clock

function

only for pH 310 meter

Type Parameter

P

No.

Default

Remarks

CLr

Memory clear

P1.0

No

Retains your current settings

CAL

View pH calibration data

P2.1

––

Calibration data for 1st buffer

P2.2

––

Calibration data for 2nd buffer

P2.3

––

Calibration data for 3rd buffer

P2.4

––

Calibration data for 4th buffer

P2.5

––

Calibration data for 5th buffer

(only in pH 310 meter)

P2.6

––

Calibration data for 6th buffer (DIN only)

ELE

View electrode offset

P3.1

0.00 mV

No offset adjustment

View electrode slope

P3.2

100.0 %

No slope adjustment

View relative mV offset

P3.3

0 mV

No mV adjustment

COF

Ready indicator

P4.1

Ready On

Ready indicator on; auto endpoint off

# pH calibration points

P4.2

3

3 pH calibration points available (1 – 5 pt

range)

Calibration buffer selection

sets(pH 310 meter only)

P4.3 USA

°C or °F (pH 310 meter)

P4.4

No default

°C or °F remains as selected

CLO

Setting year, date and time

(in pH 310 meter only)

P5.0

No default

Year, date and time remain as selected

rSt

pH factory default

For pH 300 meter

For pH 310 meter

P5.0

P6.0

No


Retains your current settings